tech-file/iec-62304/post-market-surveillance-plan.md
2023-02-27 11:31:21 +01:00

170 lines
12 KiB
Markdown

# Post-Market Surveillance Plan
This plan describes product-specific post-market surveillance activities. The general process of how to do
post-market surveillance is described in SOP Post-Market Surveillance. Its outputs are saved to the
Post-Market Surveillance Report.
## Product
| Product Name | Version | Surveillance Period |
|-------------------------|---------------|----------------------------|
| *\<your product name\>* | *\<version\>* | *\<e.g. 10/2020-10/2021\>* |
## 1. General Considerations
> Note: Whatever kind of post-market surveillance activities you define for your product, make sure to map at
> minimum all of these actions required by the MDR to one activity in section 2 below.
According to Annex III section 1.1 (b) MDR, the post-market surveillance plan shall cover:
| MDR Requirement | Activity |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------|
| A proactive and systematic process to collect any information referred to in point (a). The process shall allow a correct characterisation of the performance of the devices and shall also allow a comparison to be made between the device and similar products available on the market | |
| Effective and appropriate methods and processes to assess the collected data; | |
| Suitable indicators and threshold values that shall be used in the continuous reassessment of the benefit-risk analysis and of the risk management as referred to in Section 3 of Annex I; | |
| Effective and appropriate methods and tools to investigate complaints and analyze market-related experience collected in the field; | |
| Methods and protocols to manage the events subject to the trend report as provided for in Article 88, including the methods and protocols to be used to establish any statistically significant increase in the frequency or severity of incidents as well as the observation period; | |
| Methods and protocols to communicate effectively with competent authorities, notified bodies, economic operators and users; | |
| Reference to procedures to fulfill the manufacturers obligations laid down in Articles 83, 84 and 86; | |
| Systematic procedures to identify and initiate appropriate measures including corrective actions; | |
| Effective tools to trace and identify devices for which corrective actions might be necessary; | |
| A PMCF plan as referred to in Part B of Annex XIV, or a justification as to why a PMCF is not applicable. | |
## 2. Data Collection Activities
> Note: In the "Metric / Threshold" column there's a placeholder "(define)". Replace this with the metric and
> threshold you've decided upon.
| Activity | Assigned To | Metric / Threshold | How Often? |
|--------------------------------------------------------------------------------------------------------------|------------------------------|--------------------|-------------------------------|
| Incident documentation and analysis of undesirable side effects | QMO | (define) | 1/year on Jan 1st |
| Assess feedback (customer complaints, sales feedback) | Head of Product | (define) | 1/year on Jan 1st |
| Check SOUP for new published issues | Head of Software Development | N/A | 2/year on Jan 1st and Aug 1st |
| Research data about similar products in the market | QMO | (define) | 1/year on Jan 1st |
| Conduct post-market clinical follow-up activities as planned | Head of Medical Team | | |
| Research scientific publications | Head of Product | | |
| Research updates of standards and legislation | QMO | | |
| Analyze trends, decide on necessary measures and implement them | QMO | | 1 /year |
| Update risk management file | QMO | | |
| Compile post-market clinical follow-up report | Head of Medical Team | | |
| Compile Periodic Safety Update Report | Head of Product | | |
| Upload PSUR to Eudamed database | | | |
| Compile post-market surveillance plan and post-market clinical follow-up plan for next surveillance interval | | | |
## 3. Data Collection Categories
At minimum, the information required by the process for post-market surveillance is collected. For (enter
product name), the following categories of data will be collected specifically:
### Information about other devices*
* [BfArM Field Corrective Actions][bfarm-fca]
* Keywords:
* Filter:
* Time span:
* [BfArM Recommendations][bfarm-rec]
* Keywords:
* Time span:
* [MHRA][mhra]
* Keywords:
* Filter:
* Time span:
* [Swissmedic: List of recalls and other field corrective actions][swissmedic]
* Keywords:
* Time span:
* [FDA Recalls: medical device recalls][fda-recalls]
* Keywords:
* Time span:
* [FDA Maude][fda-maude]
* Keywords:
* Time span:
* SOUP Incident Reports
* Go through SOUP list
* Research public incident reports of the last (enter time span) months
> Make sure to not just update your clinical evaluation here and enter the same keywords as for the literature review.
> Also add keywords for the equivalent / similar devices you compare (such as their brand names).
### Other information about similar devices
* Google News Search
* Filter: (define)
* Time span:
* (...)
### Specialist literature / technical databases
> Note: This chapter should analyze other publications applicable to our product which are not considered
> already (or typically) as part of the post-market clinical follow-up.
* Pubmed
* Filter: (define)
* Time span: last 12 months
* (...)
### Serious incidents of our medical device
* Incident documentation
* Was the device involved in any event that caused, directly or indirectly, death of a patient, user or
other person or to a serious deterioration in their state of health? Did the company issue a recall of
the device?
* If so, attach respective documentation (e.g. field safety notices)
* Time span:
### Non-serious incidents and undesirable side effects of our own medical device
* Feedback and customer complaints
* Filter: hazard-related feedback, usability issues
* Time span:
### Feedback we collect from our partners, users, distributors, importers
* Feedback and customer complaints
* Filter: not hazard-related feedback, feedback on performance, safety or processes
* Time span:
The overall complaint rate is deemed acceptable if:
* Less than 10% of complaints are associated with moderate severity of harm or higher
* Overall number of complaints does not exceed 10% of total active users
* (...)
> Note: customize the acceptance criteria for the overall complaint rate to your setup
### Trends
* Compile a list of trends as described in the section below.
## 4. Trend Analysis
Trend analysis is performed with a focus on undesirable side-effects and non-serious incidents. These will be
monitored if they impact the benefit-risk ratio in a negative way.
Hazards in the risk table are compared to post-market surveillance results:
* If post-market surveillance leads to the finding that the estimated **probability** was too low (= event
happened more often in post-market surveillance), actions are initiated as described in the SOP Post-Market
Surveillance.
* If post-market surveillance leads to the finding that the estimated **severity** was too low (= event
happened led to more serious harm in post-market surveillance), a CAPA is initiated.
<!-- Links -->
[bfarm-fca]: https://www.bfarm.de/SiteGlobals/Forms/Suche/EN/kundeninfo_Filtersuche_Formular_en.html?nn=4527724
[bfarm-rec]: https://www.bfarm.de/EN/MedicalDevices/RiskInformation/Recommendations/_functions/_node.html
[mhra]: https://www.gov.uk/drug-device-alerts
[swissmedic]: https://fsca.swissmedic.ch/mep/
<!-- This is a great link -->
[fda-recalls]: https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfRES/res.cfm?start_search=1&event_id=&productdescriptiontxt=&productcode=&IVDProducts=&rootCauseText=&recallstatus=&centerclassificationtypetext=&recallnumber=&postdatefrom=&postdateto=&productshortreasontxt=&firmlegalnam=&PMA_510K_Num=&pnumber=&knumber=&PAGENUM=100
[fda-maude]: https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfmaude/search.cfm
---
Template Copyright [openregulatory.com](https://openregulatory.com). See [template
license](https://openregulatory.com/template-license).
Please don't remove this notice even if you've modified contents of this template.